Medusa R Model 999
800x-_Page 86 – Medusa R Diagram

Medusa R Model 999


Product Description

Medusa R Model 999

The Medusa R Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range 30°C to 550°C.

It is suitable for use as a radiation source for infrared thermometers. The cavity is 45 x 285mm deep and suitable for use with the larger Isotech fixed point cells. Laboratory performance and low uncertainty calibrations are ensured by the combination by high emissivity and excellent temperature uniformity.

The digital temperature controller allows the block temperature to bet set to any value from 30°C to 550°C. Uniformity of the large block is ensured by using distributed heating technology.

For the smallest of uncertainties the Medusa R may be used with Isotech ITS-90 Fixed Point Cells, Gallium 29.7646°C, Indium 156.5985°C, Tin 231.928°C and Zinc 419.527°C. The cells are provided with a certificate of metal purity.

Technical Feature Value
Model 999
Temperature Range 30°C to 550.0°C
Emissivity Greater than 0.995
Stability ±0.1°C
Display Resolution 0.01°C to 99.99; 0.1°C from 100 to 550
Heating Time 45 minutes
Aperture Diameter 45mm
Cavity Depth 285mm
PC Interface Included
Power 1000 Watts typical
Voltage 100-130 or 208-240 Vac 50/60Hz
Dimensions Height 480mm Width 425mm Depth 260mm
Weight 17kg
Options Fixed Point Cells: Indium Large Primary Cell – 998-06-00A. Tin Hockey Large Primary Cell – 998-06-00B. Zinc Large Primary Cell 998-06-00C
How to order Model 999 Medusa R – State supply voltage. Please state any special calibration requirement.