Platinum Palladium Thermocouple

NPL Platinum / Palladium Thermocouple

Categories: ,

Product Description

Platinum Palladium Thermocouple

These Platinum Palladium Thermocouple (Pt/Pd) are manufactured by the National Physical Laboratory (NPL) who have developed a novel physical mechanism which mitigates the mechanical stresses caused by the different thermal expansion of Pt and Pd wires. This dispenses with the fragile coil conventionally used with this type of thermocouple.

Platinum / Palladium thermocouples are more costly than Platinum / Platinum Rhodium types but benefit from superior performance, both in terms of stability (less decalibration) and accuracy.

Gold / Platinum Thermocouples (Au/Pt) offer the best performance but are limited to a maximum temperature of 1000°C.

These Pt/Pd thermocouples offer lower drift than Platinum / Platinum Rhodium (Pt/Pt-Rh)and with higher operating temperatures than Au/Pt. They are suitable for a range of applications including transfer standards.

In addition to developing the novel structural design along with new annealing and construction procedures NPL have established the world’s first ISO 17025 (UKAS) accredited calibration services using metal-carbon eutectic fixed point cells.

Technical Feature Value
Temperature Range 0 to 1500°C
Sheath Materials Measuring Junction 4mm x 710mm fits into a removable 7mm x 700mm protective outer ceramic sheath
Reference Junction Enclosed in a Quartz sheath
Applicable Standard IEC 62460 Edition 1.0
How to order Model type: NPL Pt/Pd Thermocouple. Includes UKAS Calibration specify range from either:
How to order 0°c to 1100°C uncertainty of ±0.2°C calibrated at fixed points Zn, Ag, and Cu. 0°C to 1330°.
How to order 0°C to 1330°C. Uncertainty of ±0.2°C from 0°C to 1100°C rising linearly to ±0.55°C at 1330°C. Calibrated at Fixed Points Zn, Ag and Pd-c.
How to order 0°C to 1500°C. Uncertainty of ±0.2°C from 0°C to 1100°C rising linearly to ±0.7°C at 1500°C. Calibrated at Fixed Points Zn, Ag, Pd-c.